Probe type | Standard probe |
---|---|
Dimensions | Ø15 x 83 mm |
Weight | 10 g |
Operating range electronics | -50…100 °C / 0…100 %rh |
Accuracy w. adjustment profile "standard" | ±0.8 %RH / ±0.1 K, at 10...30 °C at 23 °C und 10, 35, 80 %RH |
Accuracy w. adjustment profile "High Precision" | ±0.5 %RH / ±0.1 K, at 10...30 °C at 23 °C and 10, 20, 30, 40, 50, 60, 70, 80, 90 %RH |
Power supply / Power consumption | 3.3…5 VDC |
Current consumption | 4.5 mA at 3.3 VDC |
Long-term stability sensor | <1 %rh / year |
Humidity sensor | Hygromer IN-1 |
Temperature sensor | Pt100 Class A |
Probe protection | Polycarbonate plastic cage, black |
Filter cartridge | Polyethylene |
Filter type | Wire mesh filter |
Response time t 63 | <15 sec. |
Allowed wind speed | 20 m/s, with filtre |
Psychrometric calculations | Dew or frost point |
Reproducibility Airchip | <0.02 %rh / 0.01 K |
Exploitation software | ROTRONIC HW4.2 |
Probe adjustment by software | 1 Point & multipoint %rh & °C |
Firmware update | With internet and HW4 software |
User information | from Software HW4 (devicemanager) |
Data processing via HW4 | with interface cables |
Device Protection by password | Yes |
Type of output signals | 2 x 0…1 VDC |
Analogue outputs scaleable by the user | Yes |
Analog output signal (standard) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
Digital output signal (standard) | UART |
Service Interface | UART |
Reverse voltage protection | Mechanically in the connector |
Max. cable length | 5 m |
Probe material | PEEK |
Enclosure material | Polycarbonate |
Housing | Polycarbonate |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |