HC2-HK25 / HC2-HK40
HC2-HK25 / HC2-HK40

HC2-HK25 / HC2-HK40

휴대용 프로브는 높은 온도의 이동 측정에 특히 적합 합니다.                                         습도와 온도를 측정하고 이슬점 / 서리점을 계산 합니다.

Applications
기후와 온도챔버, 건조기, 에어덕트

Features
• Range of application: -100…200 °C1 / 0…100 %rh
• Digital interface (UART) and scalable analog outputs, 0…1 V
• Standard output scaling: 0…1 V = -40…60 °C / 0…100 %rh
• Adjusted at 23 °C and 10, 35, 80 %rh





Probe type High-Temperature Handheld Probe
Weight 210 g / 240 g
Operating range electronics -40…100 °C / 0…100 %rh
Accuracy at 23 ±5 °C ±0.8 %rh / ±0.1 K
Accuracy w. adjustment profile "standard" at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K
Power supply / Power consumption 3.2…5 VDC ±0 % / < 4.5 mA
Long-term stability sensor <1 %rh / year
Humidity sensor Hygromer IN-1
Temperature sensor PT100 1/3 DIN Class B
Probe protection Filter carrier, brass nickelplated
Filter cartridge Wire mesh, 20...25 um
Response time t 63 <15 sec.
Allowed wind speed 20 m/s, with filtre
Psychrometric calculations Dew or frost point
Reproducibility Airchip <0.02 %rh / 0.01 K
Exploitation software ROTRONIC HW4.2
Probe adjustment by software 1 Point & multipoint %rh & °C
Firmware update With internet and HW4 software
User information from Software HW4 (devicemanager)
Data processing via HW4 with interface cables
Device Protection by password Yes
Type of output signals 2 x 0…1 VDC
Analogue outputs scaleable by the user Yes
Analog output signal (standard) 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital output signal (standard) UART
Service Interface UART
Reverse voltage protection Mechanically in the connector
Max. cable length 2 m
Probe material PEEK
Enclosure material POM handle / sensor tube: PPS, black
FDA/GAMP compatibility FDA 21 CFR Part 11 and GAMP compatible

HC2 Probe - 간략 매뉴얼_v1.0 (PDF, 459.52 KB) 2015/10/21
HC2 Probe - 데이터 시트 (PDF, 681.39 KB) 2015/10/22
HC2 Probe - 매뉴얼_v1.34 (PDF, 702.92 KB) 2015/10/21

Contact Information

* Required Fields( * 는 필수입력 사항입니다.)