Probe type | Humidity and temperature probe for hydrogen peroxide applications (H2O2) |
---|---|
Accuracy | ±0.8 %RH, ±0.1 K, at 10...30 °C |
Measuring range with integrated probe | -50…100 °C / 0…100 %rh |
Accuracy analog output | ±1 mV |
Power supply | 3.3…5 VDC |
Current consumption | 4.5 mA at 3.3 VDC |
Humidity sensor | Hygromer HH-1 |
Continuous H2O2 load | 880 ppm / 1200 mg/m3 |
Temperature sensor | PT100 1/3 DIN Class B |
Filter type | No filter |
Response time t 63 | <15 sec. |
Calculations | Dew / Frost point |
Audit Trail/ electronic records | FDA 21 CFR Part 11 and GAMP compatible |
Analog output signal (standard) | 0…1 V = 0...100 %RH 0…1 V = -40...60 °C (default setting, freely scalable via HW4 software and AC3001 cable) |
Communication interfaces | UART |
Housing | Stainless steel 1.4301 |
IP protection rating | IP65 |
CE / EMC conform | EMC Directive 2007/108/EC |
Storage and transport conditions | -50…100 °C |