HygroPalm - HP22-A
HygroPalm - HP22-A

HygroPalm - HP22-A

HygroPalm22-A는 모든 ROTRONIC HC2 프로브와 결합될 수 있습니다. 이 것은 모든  습공기 계산을 수행할 수 있으며 습도와 온도를 측정하고 추이 지표뿐만 아니라 측정 값을 HOLD 유지 기능이 있습니다.

Applications
모바일 검사 및 HAVC에서 랜덤 테스트, 제약 산업 및 빌딩 관리 시스템

Features
• Combinable with all ROTRONIC HC2 probes
• Range of application: -10…60 °C / 0…100 %rh
• All psychrometric calculations
• Adjusted at 23 °C and 10, 35, 80 %rh
• Service interface (UART)

Applications

  • 음식
  • HVAC
  • 기상
  • 종이
  • 제약




Probe typeStandard probe
Dimensions268 x 72 x 30 mm
Operating range electronics-10…60 °C / 0…100 %rh
Accuracy at 23 ±5 °CProbe dependent
Measuring range with interchangeable probe0…100 %rh and -50…200 °C
Power supply / Power consumption9 V battery, power supply 4 mA consumption without back light
Long-term stability sensor<1 %rh / year
Humidity sensorProbe dependent
Temperature sensorProbe dependent
Response time t 63Probe dependent
Allowed wind speed20 m/s, with filtre
Psychrometric calculationsAll psychrometric values
Number of probe inputs1
Calibration / adjustmentmultipoint adjustment via keypad
Initialisation timeTypically <2 sec.
Measurement intervaltypically 1 second (without calculations)
Display typeGraphic display. Display freely configurable.
Backlightwith Backlight
Battery indicatorLow Batt. Display
Exploitation softwareROTRONIC HW4.2
Menu guidanceYes
Trend indicatorYes, for models with display
Statistical functionsMin / Max, Average
Probe adjustment by softwareYes, with service cable
Probe adjustment by keypad1 point or multi-point temperature and humidity adjustment
Event loggingSystem related events (FDA, GAMP)
Firmware updateWith internet and HW4 software
User informationfrom Software HW4 (devicemanager)
Data processing via HW4Yes
Device Protection by passwordYes
Sensor diagnostics (drift, state)Programmable; factory default - Off
Audit Trail/ electronic recordsFDA 21 CFR Part 11 and GAMP compatible
Choice of Battery / rechargeable batteryYes
Service InterfaceMini USB (UART)
Service InformationCalibration interval
Reverse voltage protectionMechanically in the connector
Max. cable length5 m
Max. probe cable length10 m
Enclosure materialABS
IP protection ratingIP40
Fire protection classComplies UL94-HB
FDA/GAMP compatibilityFDA 21 CFR Part 11 and GAMP compatible
Weight210.0000

HP22 / HP22-A - CE 인증서 (PDF, 177.89 KB) 2015/10/21
HP21 / HP22-A - 간략 매뉴얼_v1.0 (PDF, 644.48 KB) 2015/10/21
HP21 / HP22 - 데이터 시트 (PDF, 698.81 KB) 2015/10/22
HP22-A - 매뉴얼_v2.13 (PDF, 457.11 KB) 2015/10/21
HP22-A - 펌웨어 v2.4 (ZIP, 107.73 KB) 2015/10/22

Contact Information

* Required Fields( * 는 필수입력 사항입니다.)