Probe type | Standard probe |
---|---|
Dimensions | 268 x 72 x 30 mm |
Operating range electronics | -10…60 °C / 0…100 %rh |
Accuracy at 23 ±5 °C | Probe dependent |
Measuring range with interchangeable probe | 0…100 %rh and -50…200 °C |
Power supply / Power consumption | 9 V battery, power supply 4 mA consumption without back light |
Long-term stability sensor | <1 %rh / year |
Humidity sensor | Probe dependent |
Temperature sensor | Probe dependent |
Response time t 63 | Probe dependent |
Allowed wind speed | 20 m/s, with filtre |
Psychrometric calculations | All psychrometric values |
Number of probe inputs | 1 |
Calibration / adjustment | multipoint adjustment via keypad |
Initialisation time | Typically <2 sec. |
Measurement interval | typically 1 second (without calculations) |
Display type | Graphic display. Display freely configurable. |
Backlight | with Backlight |
Battery indicator | Low Batt. Display |
Exploitation software | ROTRONIC HW4.2 |
Menu guidance | Yes |
Trend indicator | Yes, for models with display |
Statistical functions | Min / Max, Average |
Probe adjustment by software | Yes, with service cable |
Probe adjustment by keypad | 1 point or multi-point temperature and humidity adjustment |
Event logging | System related events (FDA, GAMP) |
Firmware update | With internet and HW4 software |
User information | from Software HW4 (devicemanager) |
Data processing via HW4 | Yes |
Device Protection by password | Yes |
Sensor diagnostics (drift, state) | Programmable; factory default - Off |
Audit Trail/ electronic records | FDA 21 CFR Part 11 and GAMP compatible |
Choice of Battery / rechargeable battery | Yes |
Service Interface | Mini USB (UART) |
Service Information | Calibration interval |
Reverse voltage protection | Mechanically in the connector |
Max. cable length | 5 m |
Max. probe cable length | 10 m |
Enclosure material | ABS |
IP protection rating | IP40 |
Fire protection class | Complies UL94-HB |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Weight | 210.0000 |