Probes | HC2-CRP; alternatively: any HygroClip2 probe |
---|---|
Parameters | Differential pressure, temperature, relative humidity, digital and analog inputs |
Dimensions | 180 x 300 x 72 mm |
Weight | 1,700 g (with humidity sensor) 1,550 g (without humidity sensor) |
Range of application | -5...60 °C (23...140 °F) / 0...100 %RH |
Accuracy at 23 ±5 °C | ±1.0 % full-scale (Pa) |
Measurement principle | Diaphragm sensor |
Measurement ranges | ±50 Pa / ±100 Pa / ±250 Pa / ±500 Pa & -100...200 °C (probe-dependent) / 0...100 %RH |
Pressure resistance | 0.7 bar (70,000 Pa) |
Pressure connections | Front: Hose connectors Ø 6 mm x 10 mm Back: Hose connectors Ø 4 mm x 10 mm |
Medium | Air and non-aggressive gases |
Accuracy analog output | ±5 mV (voltage output) ±20 μA (current output) |
Switch output | 6 solid-state relays |
Switching capacity | <50 VDC at <2 A when observing polarity <35 VAC at <1 A, irrespective of polarity |
Zero drift | Compensated (manual or automatic zero adjustment) |
Power supply / Power consumption | 20...48 VDC / 18...35 VAC <500 mA |
Electrical connections | Screw terminals on back |
Measurement interval | 1 s |
Display type | Color TFT display |
Menu Navigation | 4 optical buttons |
Firmware update | Via HW4 software |
Analog outputs | 4, freely configurable |
Analog output signal (standard) | 0 / 4...20 mA or 0...1 / 5 / 10 V |
Permissible load | >10 kΩ (voltage output) <500 Ω (current output) |
Communication interfaces | Ethernet (Modbus TCP, HW4) / RS-485 (Modbus RTU, HW4) |
Service Interface | UART |
Housing | Front: Glass Back: Stainless Steel |
IP protection rating | Front: IP65 (also with sensors removed) Back: IP20 |
Fire protection class | Complies UL94-HB |
Mounting position | Wall Mounting |
FDA / GAMP | FDA 21 CFR part 11 / GAMP5 |
CE / EMC conform | EMC Directive 2004/108/EC |